Considerations in the calibration of the P-61 suspended sediment sampler / by P. Engel.: En13-5/94-76E-PDF

"Tests were conducted in a towing tank on five P-61 sediment samplers fitted with carefully selected nozzles. Statistical analyses of the test data were conducted. It was shown that individual samplers can be calibrated with reasonable certainty but that the variability of calibrations from sampler to sampler can be unacceptably high. It was further shown that the P-61 sampler over-samples at low velocities and under-samples at medium to high velocities. Normal fabrication variances in nozzle geometry do not affect sampler performance"--Abstract.

Permanent link to this Catalogue record:
publications.gc.ca/pub?id=9.875907&sl=0

Publication information
Department/Agency Canada. Environment Canada.
National Water Research Institute (Canada). Aquatic Ecosystem Protection Branch.
National Water Research Institute (Canada)
Title Considerations in the calibration of the P-61 suspended sediment sampler / by P. Engel.
Series title NWRI contribution ; 94-76
Publication type Series - View Master Record
Language [English]
Format Electronic
Electronic document
Note(s) Page 6 has been misnumbered as page 2.
Digitized edition from print [produced by Environment and Climate Change Canada].
Includes bibliographical references (page 7).
Includes summaries and abstracts in English and French.
Publishing information Burlington, Ontario : National Water Research Institute = Institut national de recherche sur les eaux, [1994]
Author / Contributor Engel, Peter, 1938- author.