Low cost radon detector based on spark discharge technology : report of research / prepared for CMHC by George Vandrish and Lyle Davidson, Instruscience Ltd. : NH17-289/1990E-PDF
"A portable radon detector has been successfully developed based on spark discharge technology . This detector was found to have a sensitivity to radon of 0.045 cpm/pCi. Research focused on the development of a low cost, high voltage supply, counting systems, various sensors, and a radon daughter electrostatic collector. Sensors based on conventional technology were constructed, optimised for sensitivity, simplicity of fabrication and stability. Sensors based on thin foil technology failed to function as expected and research was discontinued. Testing was ongoing throughout the project"--Executive summary.
Lien permanent pour cette publication :
publications.gc.ca/pub?id=9.905926&sl=1
| Ministère/Organisme |
|
|---|---|
| Titre | Low cost radon detector based on spark discharge technology : report of research / prepared for CMHC by George Vandrish and Lyle Davidson, Instruscience Ltd. |
| Type de publication | Monographie |
| Langue | [Anglais] |
| Format | Texte numérique |
| Document électronique | |
| Note(s) |
|
| Information sur la publication |
|
| Auteur / Contributeur |
|
| Description | 1 online resource (33 pages) : illustrations, charts |
| Numéro de catalogue |
|
| Descripteurs |
Demander des formats alternatifs
Pour demander une publication dans un format alternatif, remplissez le formulaire électronique des publications du gouvernement du Canada. Utilisez le champ du formulaire «question ou commentaire» pour spécifier la publication demandée.Détails de la page
- Date de modification :