Calibration of the SIAP ME4001 current meter / by Peter Engel and C. DeZeeuw.: En36-539/78-1982E-PDF
"Owing to the initiative of Quebec Hydro engineers, comparative ratings of five Siap ME40001 type were made four institutes including the National Water Research Institute (NWRI). Because specified procedures were not used at NWRI, the tests were repeated as well as some additional work. This analysis shows that the meter may be calibrated to an accuracy of .4% at the 99% confidence level. Analysis of the data indicates that calibrations at the NWRI towing tank are least prone to error"--Summary.
Permanent link to this Catalogue record:
publications.gc.ca/pub?id=9.904789&sl=0
| Department/Agency |
|
|---|---|
| Title | Calibration of the SIAP ME4001 current meter / by Peter Engel and C. DeZeeuw. |
| Variant title |
|
| Series title |
|
| Publication type | Monograph - View Master Record |
| Language | [English] |
| Format | Digital text |
| Electronic document | |
| Note(s) |
|
| Publishing information |
|
| Author / Contributor |
|
| Description | 1 online resource (v, 26 pages, 38 unnumbered pages) : charts. |
| Catalogue number |
|