Calibration of the SIAP ME4001 current meter / by Peter Engel and C. DeZeeuw. : En36-539/78-1982E-PDF
"Owing to the initiative of Quebec Hydro engineers, comparative ratings of five Siap ME40001 type were made four institutes including the National Water Research Institute (NWRI). Because specified procedures were not used at NWRI, the tests were repeated as well as some additional work. This analysis shows that the meter may be calibrated to an accuracy of .4% at the 99% confidence level. Analysis of the data indicates that calibrations at the NWRI towing tank are least prone to error"--Summary.
Lien permanent pour cette publication :
publications.gc.ca/pub?id=9.904789&sl=1
| Ministère/Organisme |
|
|---|---|
| Titre | Calibration of the SIAP ME4001 current meter / by Peter Engel and C. DeZeeuw. |
| Variante du titre |
|
| Titre de la série |
|
| Type de publication | Monographie - Voir l'enregistrement principal |
| Langue | [Anglais] |
| Format | Texte numérique |
| Document électronique | |
| Note(s) |
|
| Information sur la publication |
|
| Auteur / Contributeur |
|
| Description | 1 online resource (v, 26 pages, 38 unnumbered pages) : charts. |
| Numéro de catalogue |
|
| Descripteurs |
Demander des formats alternatifs
Pour demander une publication dans un format alternatif, remplissez le formulaire électronique des publications du gouvernement du Canada. Utilisez le champ du formulaire «question ou commentaire» pour spécifier la publication demandée.Détails de la page
- Date de modification :