The effect of replacing the multiband transmittance sensor glass window with a plastic lucite SAR window / author: R. Desrosiers. : En13-6/1-83-02E-PDF
"It was found that after many hundred profiles cracks began to appear and spalling occurred near the edges of the glass windows used inthe NWRI multiband transmittance sensor. The glass (Schott BK 7 UVB NVS) window was replaced with a Dupont Lucite SAR (trademark) acrylic window. The structural calculations were discussed in detail in Reference 1. Reference 2 found that temperature cycling was not the likely cause of the problem"--Introduction, page 1.
Lien permanent pour cette publication :
publications.gc.ca/pub?id=9.874867&sl=1
| Ministère/Organisme |
|
|---|---|
| Titre | The effect of replacing the multiband transmittance sensor glass window with a plastic lucite SAR window / author: R. Desrosiers. |
| Titre de la série |
|
| Type de publication | Monographie - Voir l'enregistrement principal |
| Langue | [Anglais] |
| Format | Texte numérique |
| Document électronique | |
| Note(s) |
|
| Information sur la publication |
|
| Auteur / Contributeur |
|
| Description | 1 online resource (7 pages). |
| Numéro de catalogue |
|
| Numéro de catalogue du ministère | CORRESPONDENCE FILE NO: Study 372 |
| Descripteurs |
Demander des formats alternatifs
Pour demander une publication dans un format alternatif, remplissez le formulaire électronique des publications du gouvernement du Canada. Utilisez le champ du formulaire «question ou commentaire» pour spécifier la publication demandée.Détails de la page
- Date de modification :