Uncertainty in the calibration of the DH-48 suspended sediment sampler / P. Engel. : En13-5/92-18E-PDF
"Tests were conducted in the towing tank at NWRI on a DH-48 sediment sampler with carefully selected nozzles. Statistical analysis of the test data were conducted. It was shown that this type of sampler can be calibrated with a sufficient degree of repeatability and that complete calibration of individual samplers is not necessary. It was further shown that nozzles issued with each sampler can be exchanged with similar nozzles in the field and no further calibrations are required"--Abstract.